On-line and continuous analyser of dissolved chloride or on sample, if required.
HIGHER PERFORMANCES
Relative error : < + 10% or + 3 µg/l
Stability (1 week) : + 10%
Repeatability : < + 5%
OPERATION PRINCIPLE
This chloride analyser uses a high-performance specific electrode, associated to a sample conditioning system using diluted acetic acid. All the hydraulic part is integrated in an enclosure cooled and with thermostat control at 8°C. The automatic 4-point calibration is carried out by generation chlorides. The calibration results are recorded.
This calibration system takes into account :
- past calibrations,
- acceptance criteria (e.g., calibration repeatability),
- detection of any calibration system failure.
LOWER COSTS
High measuring electrode lifetime
High availability : < 2h of intervention each month
Low cost and safety conditioning agents
ERGONOMICS
On-line help
Self diagnostic
Clear dialogue
Fully automatic calibration
Built-in self-test
Dialogue and control by drop-down menus
OPTIONS
Automatic injection system of sample solutions
Stainless support chassis with wheels
By-pass flow
Measuring line flow
Sample water temperature